Issue 06/2014 Art. N0. 7001-0179
Operating Manual
APT.line™ KMF (E5.2)
Constant climate chambers
with enlarged temperature and humidity range
with program control
Model
Art. No.
KMF 115 (E5.2)
9020-0187, 9120-0187
KMF 115-UL (E5.2)
9020-0188, 9120-0188
KMF 240 (E5.2)
9020-0145, 9120-0145
9020-0219, 9120-0219
KMF 240-UL (E5.2)
9020-0182, 9120-0182
KMF 720 (E5.2)
9020-0185, 9120-0185
9020-0245, 9120-0245
KMF 720-UL (E5.2)
9020-0186, 9120-0186
BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 100
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS +7 495 988 15 16

Table of Contents

Questions and Answers

  • N
    Nicholas HillSep 8, 2025
    Why is there condensation at the walls of the inner chamber in my Binder KMF-720?
    • S
      schneidercourtneySep 8, 2025
      If you observe condensation at the walls of the inner chamber of your Binder Laboratory Equipment, it could be due to a combination of temperature/humidity values not being in the optimum range; select a combination of temperature/humidity values in the optimum range. Alternatively, the set-point may have been too long below ambient temperature, causing icing in the preheating chamber; defrost the unit.
  • J
    Jennifer PowellSep 7, 2025
    Why is the refrigerating performance low in my Binder Laboratory Equipment?
    • M
      megan26Sep 7, 2025
      If you are experiencing low or no refrigerating performance in your Binder Laboratory Equipment, it could be due to several reasons. The ambient temperature might be higher than 25 °C / 77 °F, in which case you should select a cooler place of installation. It could also be that the combination of temperature/humidity values is not in the optimum range; select a combination of temperature/humidity values in the optimum range. If the compressor is not turned on, contact BINDER service. Also, too much external heat load could be the cause, so reduce heat load.
  • L
    Lorraine HicksAug 22, 2025
    What to do if my Binder Laboratory Equipment chamber doesn’t heat up when turned on and the safety controller responds?
    • G
      gail20Aug 22, 2025
      If your Binder Laboratory Equipment unit isn't heating up when turned on and the safety controller responds, it could be due to the limit temperature being reached. The safety controller might be set too low. To fix this, let the chamber cool down and hit the RESET button of the MB1 controller. If appropriate, select a suitable limit value. If the safety controller is defective, contact BINDER service.
  • R
    rodriguezjohnSep 4, 2025
    What to do if mechanical safety device class 3.2 responds in my Binder KMF-720 Laboratory Equipment?
    • S
      scottdylanSep 4, 2025
      If the mechanical safety device class 3.2 responds in your Binder Laboratory Equipment (with option safety device class 3.3), it could be because the limit temperature has been reached. Check the setting of the temperature set-point and of safety device class 3.2. If appropriate, select a suitable limit value. If the controller or safety device is defective, contact BINDER service.
  • J
    Jose DeanSep 3, 2025
    What to do if mechanical safety device class 3.1 responds in my Binder Laboratory Equipment?
    • N
      Nicole BishopSep 3, 2025
      If the mechanical safety device class 3.1 responds in your Binder Laboratory Equipment (with option safety device class 3.3), it might be because the limit temperature has been reached. Check the setting of the temperature set-point and of safety device class 3.1. If appropriate, select a suitable limit value. Also, too much external heat load could be the cause, so reduce the heat load. If the controller is defective, contact BINDER service.
  • A
    amytranSep 2, 2025
    Why my Binder Laboratory Equipment has no function?
    • S
      Sarah RamirezSep 2, 2025
      If your Binder Laboratory Equipment unit has no function at all, start by checking the connection to the power supply. Also, verify that the power supply voltage matches the unit's requirement of 115V or 230V. Inspect the unit fuse and replace it if necessary. If the fuse responds again, contact BINDER service. A defective controller may also cause this issue, requiring you to contact BINDER service.
  • B
    Barbara HoffmanAug 31, 2025
    What to do if my Binder KMF-720 has no unit function and a dark display?
    • C
      Chloe CampbellAug 31, 2025
      If your Binder Laboratory Equipment unit shows no function with a dark display, it could be due to the display mode being in “Standby”. To resolve this, simply press any controller key. Alternatively, the main power switch might be turned off; in that case, turn on the main power switch.
  • M
    mfieldsAug 30, 2025
    Why am I getting a humidity alarm message when operating without humidity on my Binder KMF-720 Laboratory Equipment?
    • J
      Jimmy HartAug 30, 2025
      If you receive a humidity alarm message when operating without humidity (humidity switch (4) OFF) on your Binder Laboratory Equipment, it means that the humidity set-point is set to a value > 0% r.F. In Manual Mode: Enter a humidity set-point 0% r.H. In Program Mode: Enter a humidity subprogram with humidity set-point 0% r.H.
  • J
    Janice RodriguezAug 29, 2025
    How to fix low humidity and temperature accuracy in Binder KMF-720 Laboratory Equipment?
    • A
      alexander76Aug 29, 2025
      If you're experiencing low humidity and temperature accuracy with your Binder Laboratory Equipment, it might be because the fan speed has been reduced. Set the fan speed to 100%.
  • S
    Stephanie WilliamsAug 27, 2025
    Why does my Binder Laboratory Equipment have humidity fluctuation?
    • J
      Joseph Curry DDSAug 27, 2025
      If you're experiencing humidity fluctuation and the control accuracy of ± 3 % r.F. is not reached in your Binder Laboratory Equipment, it could be due to a defective door gasket; replace the door gasket. Also, opening the door very frequently can cause this issue; try to open the doors less frequently.

Summary

1. SAFETY

1.2.1 Signal Word Panel

Defines danger levels with signal words like DANGER, WARNING, CAUTION.

1.2.2 Safety Alert Symbol

Explains the use of the safety alert symbol indicating a risk of injury.

1.5 General Operating and Installation Safety

Crucial safety instructions for unit installation and operation.

1.6 Intended Use and Restrictions

Defines proper usage of the chamber and lists restricted applications.

2. UNIT DESCRIPTION

2.1 Unit Overview

Provides a general overview of the constant climate chamber.

3. DELIVERY, TRANSPORT, STORAGE, AND INSTALLATION

3.2 Safe Lifting and Transportation Guidelines

Provides guidelines for safe lifting and moving the unit.

4. INSTALLATION AND CONNECTIONS

4.3 Freshwater Supply

Methods and requirements for supplying freshwater to the unit.

4.4 Electrical Connection

Instructions for connecting the unit to the power supply.

5. START UP

5.1 MB1 Display Program Controller Overview

Explains the functions and buttons of the MB1 controller.

5.2 Operating Modes

Describes the Idle, Manual, and Program operating modes.

6. MB1 CONTROLLER SETTINGS

7. HISTORICAL MEASUREMENT (CHART RECORDER FUNCTION)

8. MANUAL MODE

8.1 Entering Set Point Values

How to enter temperature, humidity, and fan speed set points.

9. PROGRAM OPERATION

9.1 Menu-Based Program Entry

How to enter programs using the controller's menu system.

9.2 Entry of Temperature and Fan Speed

Steps for entering temperature and fan speed values in programs.

9.3 Entry of Humidity Values

Steps for entering humidity values in programs.

10. TEMPERATURE SAFETY DEVICES

10.1 Over Temperature Protective Device (Class 1)

Description of the internal Class 1 over-temperature protective device.

10.2 Safety Controller (Class 3.1)

Description of the Class 3.1 safety controller for over-temperature protection.

10.2.2 Setting Safety Controller Parameters

Guide to checking and setting the safety controller's type and set-point.

10.3 Temperature Safety Device Class 3.3 (Option)

Details the optional Class 3.3 safety device for over/under temperature protection.

10.3.1 Temperature Safety Device Class 3.1 Operation

Instructions for setting and checking the Class 3.1 safety device.

10.3.2 Temperature Safety Device Class 3.2 Operation

Instructions for setting and checking the Class 3.2 safety device.

11. NOTIFICATION AND ALARM FUNCTIONS

11.1 Alarm System Overview

Overview of notification and alarm messages and their indications.

11.2 Resetting Alarms

Procedure for acknowledging and resetting notifications and alarms.

12. HUMIDITY SYSTEM

12.1 Humidifying and Dehumidifying System Function

Explains how the humidifying and dehumidifying systems operate.

15. MAINTENANCE, CLEANING, AND SERVICE

15.1 Maintenance Intervals and Service

Guidelines for regular maintenance and service intervals.

15.2 Cleaning and Decontamination

Procedures for cleaning and decontaminating the unit.

15.2.1 Cleaning Instructions

Detailed steps and recommended agents for cleaning the unit.

15.2.2 Decontamination Procedures

Steps for decontaminating the chamber after hazardous substance contact.

17. TROUBLESHOOTING

Heating Faults and Solutions

Addresses issues with chamber heating and safety controller response.

Refrigeration and Humidity Issues

Covers problems with cooling performance and humidity control.

Controller Malfunctions

Explains errors related to the MB1 controller display and functions.

18. TECHNICAL DESCRIPTION

18.4 KMF Technical Data

Comprehensive technical specifications for various KMF models.

Binder KMF-720 Specifications

General IconGeneral
Internal Volume720 liters
Max. Load per Shelf30 kg
Permitted Total Load120 kg
Rated Voltage230 V
Power SupplyAC
Internal Dimensions1200 x 600 x 1000 mm
External Dimensions1265 x 1985 x 985 mm
Weight258 kg