Manual for VMP300-based instruments
39
Fig. 48: Loading of the flash.
5) Click on the “Browse” button to select to the flash file (EC0368.flash) of the older
version (here: V10.12). Click on “Open” when the flash file is selected.
NOTE:
if the wrong file is selected, the communication board may be damaged.
Fig. 49: Flash location in the EC-Lab folder.
6) The pathway of
the flash of the old
version (V10.12) is
now shown in the
“Firmware
Upgrade/Downgra
de” window.
Fig. 50: “Firmware Upgrade/Downgrade” window.
7) Click on the “Load”
button. A warning
message shows
up. Click on “OK”
to continue the
downgrade.
Fig. 51: Warning message.