5. Once the drop delay calculation has been completed, click in the checkmark
button to accept the drop delay. The QC procedure estimates the optimal drop delay
and performs the side and forward-scatter detectors calibration
6. A QC report window will appear showing the QC test results. If all parameters
have passed the QC test, click in the checkmark button to close the window
7. The system is ready to acquire and sort samples