6 Safety
Failure to comply with the precautions and warnings described in this user manual may lead to
exposure to dangerous optical radiation sources. Please comply with all safety instructions and
warnings.
WARNING: Use of controls or adjustments or performance of procedures
other than those specified in this User Manual may result in hazardous radiation
exposure.
6.1 SAFETY INSTRUCTIONS
Follow these safety instructions before and during treatments:
• When the laser is in use, all operatory entrances must be marked with an appropriate warning
sign (included - see Appendix C).
• Do not operate in the presence of explosive or flammable materials. Flammable anesthetics
or oxidizing gases such as nitrous oxide (N2O) and oxygen (O2) should be avoided. Solvents
of adhesives and flammable solutions used for cleaning and disinfecting should be allowed to
evaporate before the laser is used. Attention should also be drawn to the danger of ignition of
endogenous gases.
• All persons present in the operatory must wear laser protective eyewear. Always ensure
that the laser protective eyewear is appropriate for the laser wavelength; refer to the markings
on the eyewear itself for the specifications of the protection provided.
• Prior to use inspect the eyewear for pitting and/or cracking; do not use if damaged; for
replacement or additional laser protective eyewear, please contact BIOLASE.
• Do not look directly into the beam or at specular reflections.
• Never direct or point the beam at a person’s eyes.
• Do not aim the laser at metallic or reflective surfaces, such as surgical instruments or dental
mirrors. If aimed directly at these surfaces the laser beam will reflect, creating a potential
hazard
• Always place the system into Standby mode (by pressing the control button while in Ready
mode) before exchanging Handpieces or disposable Tips.
• Always ensure that the proper laser parameters are set before the Epic X laser is used in a
clinical procedure.
• Toggle the ON/OFF switch (located on the rear of the console) to the OFF (O) position before
leaving the unit unattended.
• Do not use the Epic X if the Fiber Optic Cable is damaged or broken.
• Do not use the laser if there's a suspicion it is not functioning properly or other than
described in this User Manual.