22 Epic X User Manual
7 Installation and Set-Up
7.5 CONNECTING THE WHITENING
HANDPIECE (OPTIONAL
ACCESSORY)
The Whitening Handpiece is a reusable accessory and is
provided with disposable non-sterile protective shields
designed for single patient use. The Handpiece is non-
sterile and requires disinfection before and after each
patient treatment. This Handpiece cannot be sterilized
in the autoclave. For instructions on how to clean and
disinfect the Handpiece, refer to Section 9.
Always wipe the disposable shield with alcohol prior to use.
The disposable shield is for single-use only to avoid cross-
contamination. Discard when the treatment session is
completed.
To connect the Handpiece to the Fiber Optic Cable, push
the Handpiece onto the fiber shaft until it clicks on and is
secured.
To disconnect the Handpiece from the Fiber Optic Cable,
press and hold both buttons at the base of the fiber shaft
and gently pull the Handpiece to separate.
NOTE: Refer to the individual Instructions for Use included with the Whitening Handpiece packaging
for additional details.
7.6 CONNECTING THE DEEP TISSUE HANDPIECE (OPTIONAL
ACCESSORY)
The Deep Tissue Handpiece is a reusable accessory and is provided with disposable non-sterile
protective shields designed for single patient use. The Handpiece is non-sterile and requires
disinfection before and after each patient treatment. This Handpiece cannot be sterilized in the
autoclave, as doing so will damage its internal optics. For instructions on how to clean and
disinfect the Handpiece, refer to Section 9.
Always wipe the disposable shield with alcohol prior to use. These shields are for single-use only to
avoid cross-contamination and must be discarded after one-time use.