Instruction Manual TProfessional Basic 06/2011 57
7 Short manual
Safety Warnings
Do not open the instrument unless you are authorised to do so. Check the
label at the backside if the instrument has correct voltage configuration (110V,
115 V or 230 V).
The thermoblock and the heated lid will reach high temperatures during
operation. Both thermoblock and heated lid can burn you.
Rapid heating of the thermoblock can cause liquids to boil explosively. Always
wear safety goggles during operation. Close the lid before starting a program.
It is not necessary to apply oil into the opening of the block in order to improve
the heat transfer between the block and the sample tubes. If you still decide to
use oil, do not use silicon oil. Mineral oil may be used.
Ensure that both the rear and bottom ventilation slits at the rear and bottom of
the instrument are unobstructed. Insufficient ventilation can cause overheating
of the instrument.
The TProfessional Basic Control Panel
Navigation
Soft Keys
Back
LEDs on
Block
status
Start
Stop
Menu Buttons
Numpad
Navigation
Soft Keys
Back
LEDs on
Block
status
LEDs on
Block
status
Start
Stop
Menu Buttons
Numpad
Log in
After the TProfessional Basic has passed the initial self-diagnosis the login menu is displayed:
 Press [Log in] and log in as existing user
 Press [Supervisor Log in] and log in as supervisor
 Press [New User] and create new user account
The language displayed in the user interface can be selected (English or German). The selected
language will be stored for each individual user.