FC1400 User Manual
FC1400OM-2.04 85 / 95
FC1400 Operation Manual
Guidance and manufacturers declaration - electromagnetic immunity
The FC1400 system is intended for use in the electromagnetic environment specified below.
The customer or the user of the FC1400 system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance le
vel
Electromagnetic environment -guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
150 kHz to 80
MHz
FC1400 system must be used only in a shielded lo
cation with a minimum RF shielding effectiveness
and, for each cable that enters the shielded locatio
n with a minimum RF shielding effectiveness and, f
or each cable that enters the shielded location
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.
5 GHz
3 V/m
80.0 MHz to 2.
5 GHz
Field strengths outside the shielded location from fi
xed RF transmitters, as determined by an electrom
agnetic site survey, should be less than 3V/m.a
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note 1) These guidelines may not apply in all situations. Electromagnetic propagation is affected by a
bsorption and reflection from structures, objects and people.
Note 2) It is essential that the actual shielding effectiveness and filter attenuation of the shielded locat
ion be verified to assure that they meet the minimum specification.
a- Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone
s and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pr
edicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans
mitters, an electromagnetic site survey should be considered. If the measured field strength outside t
he shielded location in which the EUT is used exceeds 3V/m, the EUT should be observed to verify n
ormal operation.
If abnormal performance is observed, additional measures may be necessary, such as relocating the
EUT or using a shielded location with a higher RF shielding effectiveness and filter attenuation.