Rev A N-29
Purpose
T
he Triage MeterPro will prompt the user to install a Code Chip module when required.
H
owever, the
INSTALL CODE CHIP f
unction may be used as an alternate method to transfer
i
nformation from a Code Chip module into the meter’s memory. The meter will direct the user
t
o install a Code Chip module if it is attempting to run a test and does not have the data in
i
ts memory.
Types of Code Chip modules
There are five types of Code Chip modules
• Reagent Code Chip module – included in each box
of test devices. Lot number begins with a W.
• QC Sample Code Chip module – included in each box
of controls. Lot number begins with a C.
• QC Device Code Chip module – included in the black
QC Device box (comes with the Triage MeterPro) and
is labeled with the serial number of the QC Device.
•
Supervisor Access Code Chip module – included with
each Triage
®
MeterPro. Lot Number begins with SPR.
•
Software Upgrade Code Chip module – sent
separately when Triage MeterPro software
upgrades are available.
CAUTION: Running tests with the Supervisor Code Chip module installed disables all
QC Lockouts. Remove the Supervisor Code Chip module prior to running tests.
Install Code Chip
Software
V
ersion Number