Rev A N-17
PR
INTER
CO
NTRAST
T
his setting will adjust the printer contrast for meters with a serial number starting
w
ith 29901.
1
. From a list of programmable parameters, choose
ME
TER
SE
TTINGS
u
sing keys.
2. Press the key.
3. Use the keys to select
PR
INTER
CO
NTRAST
.
4
. Press the key.
5. Use the keys to select the number to the right of
CO
NTRAST
.
6. Use the keys to change the contrast of the meter’s screen.
The range is from 0 – 8 (8 being the greatest contrast).
7. Press the key to save changes.
COMMUNICATION SETTINGS
Prior to activating the LIS feature, obtain the meter’s serial number located underneath
the meter and call Biosite Technical Services at 1-888-BIOSITE (1-888-246-7483).
The Biosite Technical Services Representative will provide you with a password that, when
entered, will allow data to be transmitted.
BAUD RATE
1. From a list of programmable parameters, choose METER SETTINGS
using keys.
2. Press the key.
3.
Use the keys to select
COMM SETTINGS.
4. Press the key.
5. Use the keys to select
BAUD RATE.
6.
Use the keys to
choose 9600 or 38400.
7. Press the key.