O-10 Rev A
Purpose
R
ecall Results is used to retrieve results stored in the Triage MeterPro’s memory.
F
unction
T
he Recall Results function can retrieve these results:
•
Last Record • QC Results
•
Patient Results
–
Internal QC Results
–
Patient ID – QC Sample Results
– Panel Type – QC Device Results
– Test Device Lot Number (L/N) – Reagent Lots – QC
– User ID – Archive Last Test
– From Date • Print all Results
– Till Date – Patient Results
– QC Sample Results
– QC Device Results
• Upload Results to LIS
IMPORTANT NOTE: To safeguard the User ID Lockout, displayed and printed
User ID numbers are partially masked. The Supervisor Code Chip module is
required to be installed to view the full User ID.
LAST RECORD
Purpose: Last Record is used to retrieve the test results for the last test run.
To retrieve the last record:
1. Press the key to power The Triage MeterPro on.
2. From the
MAIN MENU select RECALL RESULTS using keys.
3. Press the key.
4. Select
LAST RECORD using the keys.
5. Press the key.
6. The results from the last test run appear on the screen.
7. Use to view the entire patient record.
8. Press the key to make a printed copy.
Recall Results
MAIN MENU
RECALL RESULTS
Note: If a result
was rejected by
the user, an
“R” will
a
ppear next to the
recalled result.