O-10 Rev A
Purpose
R
ecall Results is used to retrieve results stored in the Triage MeterPro’s memory.
F
unction
T
he Recall Results function can retrieve these results:
Last Record QC Results
Patient Results
Internal QC Results
Patient ID QC Sample Results
Panel Type QC Device Results
Test Device Lot Number (L/N) Reagent Lots – QC
User ID Archive Last Test
From Date Print all Results
Till Date Patient Results
QC Sample Results
QC Device Results
Upload Results to LIS
IMPORTANT NOTE: To safeguard the User ID Lockout, displayed and printed
User ID numbers are partially masked. The Supervisor Code Chip module is
required to be installed to view the full User ID.
LAST RECORD
Purpose: Last Record is used to retrieve the test results for the last test run.
To retrieve the last record:
1. Press the key to power The Triage MeterPro on.
2. From the
MAIN MENU select RECALL RESULTS using keys.
3. Press the key.
4. Select
LAST RECORD using the keys.
5. Press the key.
6. The results from the last test run appear on the screen.
7. Use to view the entire patient record.
8. Press the key to make a printed copy.
Recall Results
MAIN MENU
RECALL RESULTS
Note: If a result
was rejected by
the user, an
“R” will
a
ppear next to the
recalled result.