O-2 Rev A
Purpose
R
un Test is used for the analysis of a test.
O
ptions
T
he Run test function allows you to test:
•
QC Device
•
QC Sample
•
Patient Sample
In order to run a patient sample, the meter requires:
1. A valid User ID has been entered (unless this feature is bypassed or a Supervisor
Code Chip module is installed). See page N-21.
2. A QC Device has passed within the specified time period. See page O-3.
3. A QC Sample for the test device lot has passed within the specified time
period. See page O-5.
If any of these conditions exist, the meter will not continue and will prompt the user to
perform the required steps.
IMPORTANT: The Triage MeterPro will alert the user if QC has not been performed
for the device lot being tested.
Run Test
Note: The
Supervisor
Code Chip module
provides access to
functions that may
override these
Lockouts.
Note: If the
Triage MeterPro’s
internal quality control
(QC) results are
unacceptable,
exclamation points (!)
will be displayed on the
screen in the place
of the patient results.
T
o obtain r
esults, repeat
the test using a new
device.