Purpose
T
he Delete function is used to remove patient test information from the Triage MeterPro
m
emory. This becomes necessary when the meter memory becomes full.
Triage MeterPro Memory
W
ithin the Triage MeterPro, Patient Result memory can hold 750 data sets, QC Sample
c
an hold 200 data sets and QC Device memory can hold 250 data sets. When the
m
emory is full, the oldest result is automatically overwritten when a new result is added.
The meter display alerts the user as the memory becomes full.
IMPORTANT: The Delete function permanently removes all patient data from the
meter memory. Before continuing, you may wish to first print results (or upload
to your data management system/LIS).
QC Results will not be deleted.
Delete Results
Note: The
delete function
is available only when
the Supervisor Code
C
hip is inserted in
the Triage MeterPro
Code Chip module
Port.
The Memory Alert Screen