N-26 Rev A
TE
ST
SE
TTINGS
1
. From the list of programmable parameters, choose
TE
ST
SE
TTINGS
u
sing keys.
2. Press the key.
3. Select the test panel type using the keys.
4. Select the desired mode using the keys.
•
AC
TIVE
(AC
TIV
on Screen) – The test will run on every patient specimen.
• IN
ACTIVE
(IN
ACT
on Screen) – The test will never run on a patient specimen and
is dropped from all displays and printouts.
•
US
ER
IN
ACTIVE
(US
R
I on Screen) – At the time of patient testing, the user has
the ability to select the test.
• US
ER
AC
TIVE
(US
R
A on Screen) – At the time of patient testing, the user has
the ability to deselect the test.
•
QU
ALITAT
IVE
(QU
AL
), QU
ANTITAT
IVE
(QU
ANT
) or
SEMI-QUANTITATIVE (SEMI) using the keys.
5. Press the key to save changes.
6. To print a list, press the key.
QC TESTS WITH TEST SETTINGS ACTIVATED
Any test which the Supervisor has set to ACTIVE, USER ACTIVE or USER INACTIVE will run
on the QC Sample panel. The meter will display and print the test’s QC sample result, and
update the QC Sample timer for the test. If the Supervisor has set the test to
INACTIVE, the
test will have the QC Sample turned OFF. The QC Sample result won’t show for that test
and the QC Sample timer won’t be updated for it.
If the Supervisor changes the test from
INACTIVE to ACTIVE, to USER ACTIVE or to USER
INACTIVE that test may be out of QC Sample date. As always for a test with an expired QC
Sample, a passing QC Sample test will need to run before a patient sample can report a
r
esult for that test.
MAIN MENU
SET PARAMETERS
Note: When
the analyte is
deselected, the value
is not calculated nor
stored.