ome institutions may desire patient test results to be verified by the user before printing
o
r sending to the LIS. Activation of this feature displays the result, then requires the user
t
o accept or reject the result before the record is printed and saved. If the test is rejected,
t
he meter will permanently flag it as rejected by the user.
1
. From a list of programmable parameters, choose
METER SETTINGS
u
sing keys.
2. Press the key.
3. Use the keys to select
COMM SETTINGS.
4. Press the key.
5. Use the keys to select
PAT. RESULT APPROVAL.
6.
Use the keys to
choose DISABLED or ENABLED.
7. Press the key.
When enabled, after a patient result is available and displayed on the meter screen, the
user must accept or reject the result before printing, uploading or performing any other
meter functions.
The meter prompts the user to
PRESS 1 TO ACCEPT OR 0 TO REJECT. Once the test has
been accepted or rejected, the meter operates normally.
If the test has been rejected, the display and printout will be flagged with PAT. RESULT
REJECTED or RESULTS REJECTED BY USER. Recalled r
esults will display an
R befor
e
the patient ID.
MAIN MENU
SET PARAMETERS
METER SETTINGS
COMM SETTINGS