23 Implantation
Auto-initialization Auto-initialization begins automatically once the first connected lead is detected.
Auto-initialization is terminated 10 minutes after connection of the first lead. If no
other program has been transferred in the meantime, the implant subsequently
works with active automatic functions in the standard program.
Manual setting of the lead polarity or measurement of lead impedances is not
necessary.
Behavior during
auto-initialization
During reprogramming: auto-initialization is canceled and the transferred
program is immediately active.
During testing: auto-initialization is subsequently continued.
During transmission of a permanent program: auto-initialization is terminated
and the transferred program is active.
Note: After auto-initialization, all parameters are activated as in the standard
program with the following exceptions:
DDD-CLS
VVI
The automatically determined lead configuration (unipolar or bipolar) is set.