100
! Diode Test/Continuity
Rate Maximum Reading Resolution
S 1.19999V
10µ
µµ
µV
M 2.4999V
100µ
µµ
µV
F 2.499V 1mV
•
••
• Open Circuit Voltage: < +5.0V dc
•
••
• Test Current: Approximately 0.5mA dc
•
••
• Audible Tone: Continuous beep for continuity and single tone for normal forward-biased
diode or semiconductor junction
•
••
• Continuity level: Approximately below +50mVdc
•
••
• Maximum Input Protection: 500V dc or ac rms
!
Resistance/Continuity (2-wire)
Rate Range Resolution Maximum Reading Accuracy
S
120Ω
ΩΩ
Ω 1mΩ
ΩΩ
Ω
119.999
0.1% + 8
(1)
M
400Ω
ΩΩ
Ω 10mΩ
ΩΩ
Ω
399.99
0.1% + 5
(1)
F
400Ω
ΩΩ
Ω 100mΩ
ΩΩ
Ω
399.9
0.1% + 2
(1)
(1)
Use relative (REL) modifier
•
••
• Open Circuit Voltage: < +5.0V dc
•
••
• Test Current: Approximately 0.5mA dc
•
••
• Audible Tone: Continuous beep for reading is less than 10Ω
ΩΩ
Ω
•
••
• Zeroing error: 0.05Ω
ΩΩ
Ω or less (excluding test lead resistances) in each range when REL
modifier is used
•
••
• Maximum Input Protection: 500V dc or ac rms