Table 6-9. Descriptions for SR Command
Syntax Description
SR<s><nnnnnn>
SR command is used to set the relative base for relative function.
<s> is a sign symbol “+” or “-“.
<nnnnnn> is a six-digit number from “000000” to “199999”.
Example: “SR+001000”
Rate Range Relative base modifier to be
Slow 120.000 V +1.000V
Medium 400.00V +10.00V
Fast *1 400.0V +10.0 V
Notes:
The least setting digit is blank on the display of meter, but it still uses to relative base. To set least
setting digit to “0” for fast mode as necessary.
Table 6-10. Descriptions for SO Command
Syntax Description
SO<nn> command is used to select the reference impedance for dBm
calculation.
<nn> is a two-digit decimal numeric number from “00” to “20”,
representing 21 different types of reference impedance.
Example: Command string “SO15” to set reference impedance at 600.
nn Impedance nn Impedance nn Impedance
00 2 07 110 14 500
01 4 08 124 15 600
02 8 09 125 16 800
03 16 10 135 17 900
04 50 11 150 18 1000
05 75 12 250 19 1200
SO<nn>
06
93 13 300 20 8000