! DC Current
Rate Range Resolution
Full Scale
Reading
Accuracy (1 year)
Burden Voltage
(1)
& Shunt Resistor
S
12mA
120mA
1200mA
12A
0.1µ
µµ
µA
1µ
µµ
µA
10µ
µµ
µA
100µ
µµ
µA
11.9999
119.999
1199.99
11.9999
0.05% + 15
(2)
0.05% + 5
0.15% + 5
0.2% + 5
<0.15V / 10Ω
ΩΩ
Ω
<1.5V / 10Ω
ΩΩ
Ω
<0.3V / 0.1Ω
ΩΩ
Ω
<0.6V / 0.01Ω
ΩΩ
Ω
M
40mA
120mA
1200mA
12A
1µ
µµ
µA
10µ
µµ
µA
100µ
µµ
µA
1mA
39.999
119.99
1199.9
11.999
0.1% + 6
0.1% + 3
0.15% + 3
0.2% + 3
<0.5V / 10Ω
ΩΩ
Ω
<1.5V / 10Ω
ΩΩ
Ω
<0.3V / 0.1Ω
ΩΩ
Ω
<0.6V / 0.01Ω
ΩΩ
Ω
F
40mA
120mA
1200mA
12A
10µ
µµ
µA
100µ
µµ
µA
1mA
10mA
39.99
119.9
1199
11.99
0.1% + 2
0.1% + 2
0.15% + 2
0.2% + 2
<0.5V / 10Ω
ΩΩ
Ω
<1.5V / 10Ω
ΩΩ
Ω
<0.3V / 0.1Ω
ΩΩ
Ω
<0.6V / 0.01Ω
ΩΩ
Ω
(1)
Typical at full scale reading and voltage across the input terminals
(2)
Use relative (REL) modifier
•
••
• Maximum Input and Overload Current Protection (for dc and ac current):
mA Input Terminal: 1200mA dc or ac rms and protected with 2A/250V, IEC-127 sheet 1 fast
blow fuse;
12A input terminal: 10A dc or ac rms continuous, or 12A dc or ac rms for 30 seconds
maximum and protected with 15A/500V, breaking capacity 10,000A fast blow fuse.
•
••
• Response Time: Approximately 1.0 second when the displayed reading reaches 99.9% dc
value of the tested input signal at the same range.