SYSTem Subsystem : ( SCPI approved ).
SYSTem:VERSion?
Description : Return the firmware version of the meter.
SYSTem:ERRor?
Description : Return the next message from the system error queue.
Response : <string> format, Table 7-5 is a list of SCPI error message that might occur
during operation.
Table 7-5. SCPI Error Message
Number Error String Number Error String
0 No error -108 Parameter not allowed
-100 Command error -109 Missing parameter
-102 Syntax error -200 Execution error
-103 Invalid separator -222 Data out of range
-104 Data type error -224 Illegal parameter value