29
CAUTION
Users must use IT-E131 to realize the remote operation between PC and 8500
series electronic load.
Turn-On Checkout
Introduction
Successful tests in this chapter provide a high degree of confidence that the electronic
load is operating properly.
Checkout Procedure
The test in this section checks for proper operation of the electronic load. If you have
not already done so, connect the power cord to the unit and plug it in.
Procedure Display Explanation
1. Turn the unit on.
The electronic Load
undergoes a self-test
when you First turn it on.
During self test, all
segments are briefly
lit
2. Wait for 1s after turn
on electronic load.
EPROM ERROR
EEPROM damage or
Lost data of last
power off
Run well if no such
display, system will
go to the step 3
directly.
3. Wait for another 2S.
ERROR CAL.DATA
EEPROM Lost
calibration data
Run well if no such
display, system will
go to the step 4
directly.