SPECIFICATIONS
Below are some remarks in regards to all specifications pertaining to the 880 LCR meter.
*Specifications are subject to change without notice.
Notes:
1. Measurement performed at the test socket.
2. Measurements performed after correct open and short calibration.
3. DUT and test leads must be properly shielded to guard if necessary.
4. Based after 30 minutes of warm up time and operated at 23 °C + 5 °C, <75% R.H.
5. Q value is the reciprocal of DF.
6. Accuracies based within 10% to 100% of full scale of range; values outside of range
should be used as reference only.
7. Based on battery powered operation.
8. --- means parallel or series measurement mode.
General Specifications
Accuracy is 0.02% of
actual frequency
100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz(test setting)
100Hz, 120.048Hz, 1 kHz, 10 kHz (actual frequency)
Test Signal Level (typical)
0.3 Vrms, 0.6 Vrms, 1 Vrms.
Series mode / Parallel mode
0.1% (See accuracy specifications for details).
3-terminal, 5-terminal with sockets
Max. Counts of Primary Parameter: 40,000; D / Q / θ minimum
resolution of secondary parameter: 0.0001.
16 Hours (typical) based on backlight off and new alkaline
6 Hours (typical) based on backlight off and new fully charged Ni-
MH battery