49
Command Reference
*IDN?
Query instrument ID.
Return: <instrument model>, <firmware version>, <serial number>
*LLO
Local Lockout. This means that all front panel buttons, including the RMT key is not available.
(POWER button is enabled.)
*GTL
Go to local and remove local lockout. If *LLO is sent, the only way to operate front panel is to
go to *GTL.
*TRG
Trigger the instrument to take a measurement. Due to the automatically continuous test, *TRG
command is of no use.
SCPI Commands
This section described all the SCPI commands supported by the meter. The meter can accept
both upper case and lower case commands.
Text Symbol
Meaning
[ ]
Option; can be omitted
|
Exclusive OR
< >
Defined element
( )
Comment
?
Question mark
:
Separated two command keywords
Table 6 - SCPI Symbol Conventions
FREQuency Subsystem commands
FREQuency <value>
Description: Set the test frequency
Parameters: 100, 120, 1000, 10000,100000 or 100Hz,120Hz,1kHz,10kHz,100kHz (model
supported)
Example: FREQuency 100
Set the frequency to 100Hz
FREQuency?
Description: Query the current test frequency
Return: <100Hz|120Hz|1kHz|10kHz|100kHz>