37
Parameter: 0~255
Reset value: consult PSC command
Example: STAT:OPER:ENAB 110
STATus:OPERation:ENABle?
This command queries parameters from the operation enable register
Command syntax: STATus:OPERation:ENABle?
Return parameter: <parameter of operation enable register>
Example: STAT:OPER:ENAB?
STATus:OPERation?
This command queries the operation condition register. After this command executed, the operation
condition is reset.
Command syntax: STATus:OPERation[:EVENt]?
Return parameter: <parameter of operation event register>
Example: STAT:OPER?
STATus:OPERation:INSTrument?
This command queries the operation event sub register. After this command executed, the operation event
sub register is reset.
Note: this command is only valid for current channel.
Command syntax: STATus:OPERation:INSTrument[:EVENt]?
Return parameter: <parameter of operation event sub register>
Example: STAT:OPER:INST?
STATus:OPERation:INSTrument:ENABle
This command sets the parameter of operation event enable sub register. Setting parameter can determine
which bit value of operation event enable sub register is 1 and the bit will enable OPER of status byte
register is 1.
Note: this command is only valid for current channel.
Command syntax: STATus:OPERation:INSTrument:ENABle <value>
Parameter: 0~255
Reset value: consult PSC command
Example: STAT:OPER:INST:ENAB 110
STATus:OPERation:INSTrument:ENABle?
This command queries the parameter of the operation event enable sub register. After this command
executed, operation event enable sub register is reset.
Note: this command is only valid for the currently active channel.
Command syntax: STATus:OPERation:INSTrument:ENABle
Return value: <parameter of operation event enable register>
Example: STAT:OPER:INST:ENAB?