Query syntax: OUTPut:TIMer?
OUTPut:TIMer:DATA
This command is used to set the output timer’s time.
Command syntax:OUTPut:TIMer:DATA <NRf>
Parameter: 0-99999.9
Unit:S
Query syntax: OUTPut:TIMer:DATA?
Return parameter: <NRf>
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
This command is used to set the output current value. MIN represents 0 A and MAX represents
the max value of the current range. CURR? MIN or CURR? MAX commands are used to query
the minimum and maximum current of the range. DEF represents the default value and can also
be used to set the value. CURR UP and CURR DOWN commands and also be sent to increase or
decrease the current. The step value should be set with command “CURR:STEP”.
Command syntax: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf>
Parameter: MIN to MAX|MIN|MAX|UP|DOWN|DEF
Unit: A|mA|uA
Query syntax: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?; CURRent?
<MINimum|MAXimum>
Parameter: None
Return parameter: <NRf>
CURRent:STEP
This command is used to set the step value for CURR UP and CURR DOWN commands.
CURR:STEP? DEF command can query the step resolution of current for this unit. For example, a
step value of 0.01 represents 10 mA.
Command syntax: [SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement] <NRf>
Parameter: MIN to MAX
Query command: CURRent:STEP?
Return parameter: <NRf>
Example: CURR:STEP 0.01 //set the step value to be 0.01 A
CURR UP //increments output current by 0.01 A
CURR:STEP 0.02 //set the step value to be 0.02 A
CURR DOWN //decrements output current by 0.02 A
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]{<current level>|MINimum|MAXimum}
This command is used to set a current value to be triggered. This value will be set after the
instrument receives a trigger signal. Sending CURRent command will not impact this command’s
current setting value. Sending a query command will return the original setting value.
Command syntax: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <NRf>
Parameter: MIN to MAX|MIN|MAX
Unit: A