YH-600B Pro User Manual V3.0
75 / 88
Fig. 2-48
Analysis range input is yes, default entire file (light on to light
off). As the input is no, users can manually input the strat
frame and end frame. The automatic analysis range is from
start frame to end frame, and the analysis range mainly
affects the results of automatic analysis and report
generation.
The Respiratory Event includes Max. APN amplitude, Max.
HYP amplitude and Minimum duration. The setting of these
parameters will affect the results of automatic analysis of
respiratory events.
SpO
2
Event includes Min. SpO
2
Desaturation, Min. valid
SpO
2
and Max. Valid SpO
2
. The setting of these
parameters will affect the analysis results of oxygen