OZONE ANALYZER BMT 965 Manual, Rev. 01/2024
___________________________________________________________________________
30
M965_001_EN_2023
-
11
-
01
The *.965 files are used by the Windows software BMT 965 Link, which comes with the instru-
ment. Concentration logs only exist if logging is enabled. All files can be copied from the instru-
ment, e.g. for the purpose of sending them to the BMT service department for further diagnostic
analysis.
Log files can be deleted from the instrument either by selecting the corresponding front panel
menu item, or by using Windows Explorer when the BMT 965 is connected to a computer.
Remark: The files <Serial number>_Dia.965 and ID.965 and Error and Event logs should never
be deleted. They are needed for diagnostic purposes and should be sent to BMT in case of a service
issue.
The BMT 965 drive also contains a folder "Docs", which stores this manual and some technical
notes, describing practical uses of the instrument.
Another folder "Run_965Link" contains the communication software BMT 965 Link which allows
configuration of the instrument and visualization of the logs described above. To start the software,
just double click on L965.exe. The software immediately starts communication with the instrument
via USB.
The folder "Setup_965Link" contains a setup program for permanent installation of BMT 965 Link
on your computer.
The folder "Teamviewer" holds a portable version of the Teamviewer software. If the connected
computer has contact to the Internet, this software in conjunction with BMT 965 Link can be used
for remote diagnostics. The screen of the connected computer can be shared with e.g. a BMT
service technician. Please call BMT in advance to schedule a Teamviewer session.
Connecting a USB Flash Drive to the BMT 965
Only flash drives (sticks) with a FAT32 file system are supported. If you are not sure, how your
drive is formatted, check on a Windows computer: with the drive inserted, right click on it in
Windows Explorer to see the properties. You should see FAT32 as the file system. If this is not
the case, you can format the drive and choose FAT32 in the process. Be aware, that formatting will
erase all data on the drive.
Hint: Flash drives with a capacity larger than 32GB cannot be formatted to FAT32 on a Windows
computer using onboard tools, only ExFat or NTFS, which will not work on the BMT 965.
Note: While having a USB host adapter with or without a flash drive or a PC connected to the
BMT 965 logging on the SD-card is restricted.
From the accessories, take the USB host adapter and connect it to the USB flash drive. Connect
the USB host adapter to the BMT 965. The display will show
Copy files?
Pressing BACK aborts. If you press ENTER, all log files in internal storage are copied into a folder
BMT. If such a folder does not exist, it will be created. Copy progress is shown on the display. Old