© Bosch Healthcare Solutions GmbH 15Instructions for use
Menu
To configure or set up a new Ethernet connection:
Select configure network:
As network type select Ethernet 1.
Select DHCP enabled to support dynamic host configuration protocol from the network.
Select DHCP disabled to set the following components manually:
Internet protocol address (IP address)
Subnet mask
Gateway
Follow the instructions displayed on the screen.
To configure or set up a new WLAN connection:
Select configure network:
As network type select WLAN. The following components have to be set manually:
Service set identifier (SSID)
Security protocol
WLAN password
Select DHCP enabled to support dynamic host configuration protocol from the network.
Select DHCP disabled to set the following components manually:
Internet protocol address (IP address)
Subnet mask
Gateway
Follow the instructions displayed on the screen.
If necessary a proxy can be configured.
To view the current configuration, select show network configuration.
5.4 Configuration of order/result interface (HIS/LIS)
The analyser exchanges information with hospital and laboratory information systems (HIS/LIS), if connected.
Itreceives orders from the system and uploads results.
Log in as an administrator for access.
4
CONTINUE
enabled
In the menu, select settings, select network/interfaces, then select order/result interface.
To configure or set up a new order/result interface to HIS/LIS:
Select configure interface:
The following components have to be set manually:
Inbox location
Outbox location
Username
Password
Follow the instructions displayed on the screen.
To view the current configuration, select show interface configuration.
The network must be connected to set up an HIS/LIS configuration.