EasyManua.ls Logo

Bosch Worcester GC2000iW 30 C NG - Menu 6; Menu 0

Bosch Worcester GC2000iW 30 C NG
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Settings in the service menu
Worcester 2000 – 6 721 814 551 (2019/09)
50
8.2.6 Menu 6
Press the ( key and * key at the same time until L.1 appears.
Press arrow key until L.6 is displayed.
To confirm the selection: press OK key.
Selecting and setting the service function.
The basic settings are depicted as highlighted in the following table.
Table 26 Menu 6: function checks
8.2.7 Menu 0
Press the ( key and * key at the same time until L.1 appears.
Press arrow key until L.0 is displayed.
To confirm the selection: press OK key.
Selecting and setting the service function.
The basic settings are depicted as highlighted in the following table.
Table 27 Menu 0
Service function Settings/adjustment range Remark/restriction
6-t1 Permanent ignition OFF: switched off
ON: switched on
Testing the ignition by means of permanent
ignition without gas supply.
To prevent damage to the ignition
transformer: Leave the function switched on
for a maximum of 2 minutes.
6-t2 Permanent fan operation 0 … 100 % Fan running without gas supply or ignition.
6-t3 Permanent pump operation (heating
pump)
0 … 100 % If a value >0 is set, the pump runs with 100 %.
6-t5 3-way valve permanently in the DHW
heating position
0: heating
1: DHW
2: (not available)
6-tA Ionisation oscillator OFF: switched off
ON: switched on
6-tb Burner test 0 … 100 % The burner test is ended by resetting the setting
value to 0, or by exiting L.5.
Service function Settings/adjustment range Remark/restriction
0-A1 Manual operation OFF: switched off
ON: switched on
0-A2 Set temperature in manual operation 30 … 60 ... 82 °C

Table of Contents

Related product manuals