Specications and Technical Data
Clinician Manual
97035873-01 61 of 75
Specications and Technical Data
Parameter Range Default
Areas (Channels) 4
Amplitude 0 – 20 mAa 0 mA
Rate 2 – 1200 pps
b
40 pps
Width 0 – 1000 µsec 210 µsec
Cycle 0 – 90 min, OFF OFF
Ramp ON 1 – 10 secs 3 secs
Contacts 1 – 16; +, -, OFF 1 – 16: OFF
a. The Precision System includes programmable coverage areas with each individ ual electrode contact limited to 12.7 mA. A
programming interlock is enforced to limit the coverage area output current to 20 mA or less. For example, a maximum current
output of 12.7 mA on a rst electrode would limit the total summed current output on remaining electrodes to 7.3 mA within one
coverage area.
b. Only one Area is available if the rate is 130 pps.
Materials
The Stimulator contains a radiopaque identication tag. The identication tag is visible using standard x-ray
procedures.
Case Titanium
Header Epoxy
Strain Relief Silicone
Size/Volume 55 mm x 45 mm x 11 mm / 20.7 cm
3
(including header)