Electromagnetic Compatibility
Clinician Manual
97035873-01 67 of 75
Electromagnetic Compatibility
EN 60601-1-2 Classication Information
• Internally Powered Equipment
• Continuous Operation
• Ordinary Equipment
• Class II
The Precision System is intended for use in electromagnetic environment specied below.
The customer or the user of the Precision System should assure that it is used in such an
environment.
Emissions Test Compliance Electromagnetic Environment-guidance
RF emissions CISPR 11 Group 1 The Precision System uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 Class B The Precision System is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage uctuations / Flicker
emissions IEC 61000-3-3
Not Applicable