100-214-239 Rev. A 4-17
Guards and Safety Equipment
Digital Sonifier
4.7 Guards and Safety Equipment
Although the Digital Sonifier operates outside the normal range of human hearing, some appli-
cations can create audible noise above 85dB. If an uncomfortable level of noise is present, the
operator should wear ear protection for saf e operation.
Appropriate eye protection should be worn when operating the Digital Sonifier, to prevent
possible splash injury originating in the solution.
WARNING
j The Ultrasonic Horn can cause injury and/or equipment damage during oper-
ation. To avoid injury or accident, never touch the Ultrasonic Horn while the
System is turned on, and do not allow the Horn to come in contact with solid
vessels or supports.
The User I/O may be used to remotely control the System. If this is the case, you must design
in whatever safety precautions are appropriate to your User I/O circuit design to prevent unex-
pected sta rt- up, whi ch can cau s e perso na l inj ury and can caus e equ ipm ent dam ag e.