8 (41) BRUKER BCU05 Version 006
BCU05 HR applications
Technical specifications 1.1
Sample temperature range -5°C to + 50°C
Input :
Taken from BVT3X00
• Dry air (dew point < - 50°C) or dry N
2
• Pressure 0,3 bar
• Flow Rate 300-2000 l/h (Probe dependant)
Ambient temperature range 20 to 32 °C
Mains connection 1x 208/220/240 Vac - 50/60 Hz - 450 Watts
Warm-up time 1/2 hour
Case dimensions 550 x 500 x 490 mm
Trunk lenght
• BCU05 3 meters
• BCU05 LT 4 meters
Weight
• BCU05 35 kg
• BCU05 LT 46 kg
Gas content 0,4 kg Freon R404A
Output
Cooled air or N
2
with temperature reduced by approx. 65°C.
Stability of output temperature
± 1°C under perfectly constant conditions of mains voltage, room temperature, air
flow, etc.
Operating principle
The BCU05 is a cooling unit that reduces the temperature of an air input by typi-
cally 65°C.
Freon gas is circulating around an external heat exchanger. The system is com-
pletely sealed. The unit is designed to be operated in conjunction with the or
BVT3X00. The VTU provides the gas input for the BCU05 as well as controlling
the sample temperature. When operated in conjunction with a BTO2000, sample
temperature stability of ± 0.01°C/°C can be achieved (with BVT3X00).