EasyManua.ls Logo

BTX Gemini X2 - Experiment Troubleshooting; Arcing; Low (or no) Transfection Efficiency, or Incorporation; Low Viability

BTX Gemini X2
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gemini Series Electroporator User’s Manual
toll free 800.272.2775 • 508.893.8999 • www.btxonline.com
45
Experiment Troubleshooting
Arcing
Verify electrical component functionality. Verify properties of cell
sample (do cells need to be washed? Is the buffer appropriate for
application?). Verify properties of transfectant molecule (Is the
DNA well purified?) Try reducing the voltage or increase sample
volume until arcing is no longer a problem.
Low (or no) Transfection Efficiency, or
Incorporation
(Verify physical, biological, chemical parameters) Verify delivery
of the pulse and pulse parameters. Is the voltage correct? Chamber
gap? Pulse length or appropriate instrument settings? Number of
pulses? If so, follow Optimization Guidelines outlined.
Low Viability
Verify physical, biological, chemical parameters. Is the voltage,
chamber gap, pulse length (time constant), pulse number and
other instrument settings correct? If so, reduce voltage, pulse
length, or number of pulses and re-optimize protocol to improve
viability as outlined .
Voltage Drop
A drop in output voltage accompanies pulse delivery into highly
conductive samples (for example, PBS). Thus the displayed voltage
may in these situations be less than that expected, given 5% full
scale accuracy and the monitoring accuracy of 5%.

Table of Contents

Other manuals for BTX Gemini X2