EasyManua.ls Logo

Burkert 8045 - Operating and Functions; Safety Instructions; Operating Levels of the Device

Burkert 8045
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
Operating and functions
8. OPERATING AND FUNCTIONS
8.1. Safety instructions
Warning
Risk of injury due to non-conforming operating.
Non-conforming operating could lead to injuries and damage the device and its surroundings.
The operators in charge of operating must have read and understood the contents of this manual.
In particular, observe the safety recommendations and intended use.
The device/installation must only be operated by suitably trained staff.
8.2. Operating levels of the device
The device has two operating levels: the Process level and the Configuration level.
The Process level makes it possible:
to read the flow rate measured by the device, the value of the current transmitted on the 4-20 mA analogue
output, the values of both the daily and main totalizers.
to reset the daily totalizer.
to access the Configuration level.
The Configuration level comprises three menus (Parameters, Test and Information) and makes it possible:
to set the device parameters.
to test some device parameters.
to calibrate the device.
to read, when the status LED of the device is orange or red, the warning and fault messages generated by the
device.
Table 1: Default settings of the device
Function Default value Function Default value
LANGUAGE English OUTPUT DO3 Hysteresis
UNIT of the flow rate l/min. 3-= 0.000
UNIT of the totalizers litre 3+= 0.000
K-FACTOR 1.000 Not inverted
OUTPUT AO1 4mA= 0.000
time delay = 0
20mA= 0.000 INPUT DI1 disable
OUTPUT DO1 pulse FILTER 5, slow
PU= 0.00 litre FREQUENC. 50Hz
English
Type 8045

Table of Contents

Other manuals for Burkert 8045

Related product manuals