53
Operating and functions
8.7. Details of the Test menu
To access the Test menu, simultaneously press keys
0......9
ENTER
for at least 5 s.
This menu comprises the following configurable parameters:
CAL AO1
Process level
CALiB. 0
FLOw
K-sENsOR
FLOw-w.
12.6 L/s
END
0......9
Adjusting the 4-20 mA output
sAVE N/Y
Calibrate the flow zero point of the device.
Checking the correct operating of the outputs
Setting the coefficient Kw of the flow sensor to adjust the device accuracy.
Setting the flow rate range outside which a warning message is generated.
Saving the changes made within the Test menu or
not. If the changes are saved, the device operates
with the new settings.
Fig. 54: Diagram of the Test menu