5 | Checking BYK-Gardner GmbH
26 wave-scan 3 ROBOTIC 301 200 141 E 2401
5.3 Procedure
Perform a measurement with the robot. The values for a correct reading are
printed in the retraceable certificate and on the reference tile. These values can be
entered in smart-robotic for a pass / fail check.
Illustration26: Daily check in smart-robotic
If the value measured on the reference tile is within the printed tolerance range,
the requirements are met.
If the mean value is not within the desired tolerance range, try carefully cleaning
the test tile. If this produces no improvement, please contact our customer service
department.
http://www.byk-instruments.com/contact-infos