BYK-Gardner GmbH Appendix | 6
301 200 141 E 2401 wave-scan 3 ROBOTIC 31
6.5 Technical Data
6.5.1 Measuring Data
Light source Laser Diode LED and IR-SLD
Laser diode Laser Class 1, P ≤ 390 µW, λ=670 nm,
DIN EN 60825-1:2015-07
Measure range Dullness: 1 to 40
LW & SW: 0 to 100
Wa - We: 0 to 100
Structure spectrum Dullness: < 0.1 mm
Wa: 0.1 to 0.3 mm
Wb: 0.3 to 1.0 mm
Wc: 1.0 to 3.0 mm
Wd: 3.0 to 10.0 mm
We: 10.0 to 30.0 mm
Scan length 5 / 10 / 20 cm
Scan speed 50 to 150 mm/sec
Repeatability du < 40: 4% or > 0.4 (standard deviation)
du > 40: 6% or > 0.6 (standard deviation)
Reproducibility du < 40: 6% or > 0.6 (standard deviation)
du > 40: 8% or > 0.8 (standard deviation)
Distance to surface 15 ± 2mm
Angle to perpendicular ± 2°
Object radius > 500 mm
Smallest sample size 35 mm x 150 mm
Resolution 375 pixel per cm
Memory capacity n/a - immediate data transfer
6.5.2 General Data
Temperature range Storage: 0° C to 60° C
Operation: 10° C to 40° C
Relative humidity Up to 85% non-condensing at 35° C
Operation altitude Up to 2000 m (6561 ft)
Dimensions (LxWxH) 112 x 59 x 131 mm (4.4 x 2.4 x 4.8 in)
Weight Ca. 550 g (1.21 lbs)
Interface Initialization: USB-C (USB 2.0)
Operation: Ethernet LAN (M12-A)
Power supply (input) Initialization: 5 VDC ; max. 1.75 A (USB-C)
Operation: 24 VDC ; max. 0.75 (M12-A)