8200-M035-O1 Rev C 200 Installation & Technical
31
SETUP AND CALIBRATION, CONT.
Dual-Point without Zero (False Zero) Calibration
IMPORTANT! The functions of numeric keys are replaced by using the
UNITS/LEFT ARROW and the ASTERISK/UP ARROW keys. The cursor location is
identified by the blinking character and can be advanced to the left to the next
position by pressing the UNITS/LEFT ARROW key. Pressing the ASTERISK/UP
ARROW key will change the blinking character to the next value.
CAL1= First Calibration Weight
The display will show CAL1=0. This is the first of two calibration steps. This weight is
the TEST WEIGHTS / TEST LOAD.
Place the weights on the scale platform.
Using the UNITS/LEFT ARROW and ASTERISK/UP ARROW keys, input the value of the
calibrated test weights / test load, then press the TARE/ENTER key. NOTE: When
entering values for CAL1=, the digits start displaying on the right side of the display
and proceed to the left. When large values are used (more than 3 digits), the CAL1=
prompt will automatically scroll off the left side of the display to show the additional
digits on the right as they are entered.
Starting at the left and advancing right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: CAL2=.
CAL2= - Second Calibration Weight
The display will show CAL2=0. This is the second of two calibration steps.
Remove the weights on the scale platform, and then press the NET/GROSS key.
Starting at the left and advancing right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the display
will show: Sio?.
Single-Point for Span Only (Last Zero) Calibration
CAL1= First Calibration Weight
The display will show CAL1=0. This is the first of two calibration steps. This weight is
the TEST WEIGHTS / TEST LOAD.
Zero the scale, and then place the weights on the scale platform.
Using the UNITS/LEFT ARROW and ASTERISK/UP ARROW keys, input the value of the
calibrated test weights / test load, then press the TARE/ENTER key. NOTE: When
entering values for CAL1=, the digits start displaying on the right side of the display
and proceed to the left. When large values are used (more than 3 digits), the CAL1=
prompt will automatically scroll off the left side of the display to show the additional
digits on the right as they are entered.
Starting at the left and advancing right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: CAL2=.
CAL2= - Second Calibration Weight
The display will show CAL2=0. This is the second of two calibration steps.
Remove the weights on the scale platform, and then press the ZERO key.
The display will advance to Sio?.