8200-M035-O1 Rev C 200 Installation & Technical
CALIBRATION “C” NUMBERS
The “C” numbers are displayed only during the Test mode operation by pressing the
ASTERISK/UP ARROW key then the UNITS/TEST key. The "C" numbers are shown at the end
of the test operation and each number is displayed for approximately 4 seconds, allowing you
to record them. Each number may be up to three (3) digits in length. By recording these
numbers you will be able to return the indicator to its present calibration settings without using
test weights simply by entering the “C” numbers.
1. With CAL1= displayed, press the UNIT key.
2. At the C1= prompt, press the TARE/ENTER key to show the current value of the C1
number.
3. If the C1= number displayed is acceptable, press the TARE/ENTER key again to
save it.
4. Otherwise, use the ASTERISK/UP ARROW and UNITS/LEFT ARROW keys to enter a
new C1= number, then press the TARE/ENTER key.
5. Repeat steps 2 through 4 for C2=, C3= and C4=.
NOTE: If any components have been changed that affect calibration and/or
your scale is used in a commercial application and must be "Legal for Trade"
you cannot use the “C” numbers to re-calibrate.