8200-0753-0M Rev A 225D Weight Digital Indicator
18
COMMUNICATION ERROR BETWEEN LOAD CELLS X AND Y
Probable Cause
Items to Check
There is a loss of
communication between load
cells.
Check that cable is connected correctly.
Check cable for damage.
Verify that connector is clear of debris.
Check load cell COM ports on both load cells.
LOAD CELL X, Y, Z NOT RESPONDING
Probable Cause
Items to Check
The 225D cannot detect the
exact communication problem
with the unresponsive load
cell.
Check that cable is connected correctly.
Check cable for damage.
Verify that connector is clear of debris.
Potentially dead load cell.
LOAD CELL X DAMAGED
Probable Cause
Items to Check
There is irreparable internal
damage to the load cell.
Replace load cell.
NO WEIGHT
COMMUNICATION ERROR BETWEEN
LOAD CELLS 1 AND 3
ID COUNT
lb
G




-------
NO WEIGHT
LOAD CELL 2 NOT RESPONDING
ID COUNT
lb
G



-------
NO WEIGHT
REPLACE LOAD CELL 2
ID COUNT
lb
G



-------