EasyManuals Logo

Cardinal APS8 User Manual

Cardinal APS8
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
8527-0544-2M Rev G APS Enterprise Scale 11
SETUP AND CONFIGURATION, CONT.
UnS (Motion Range)
This is the motion range of the scale, or the number of divisions that the scale weight can
change without setting the motion indicator.
Press the UNITS key to show current setting. If value displayed is acceptable, press the
UNITS key again to save it. Otherwise, press the ZERO key to step through selections for
a new value and then press the UNITS key to save it.
Allowable values are 0 through 9.
FILtr (Adjustable Digital Filtering)
This is the adjustable digital weight filtering that is applied to the weight in order to help filter
out vibrations or other noise that might make the weight unstable.
Press the UNITS key to show current setting. If value displayed is acceptable, press the
UNITS key again to save it. Otherwise, press the ZERO key to step through selections for
a new value and then press the UNITS key to save it.
Allowable values are Off, 1, 2 or 3.
OFF
= No weight filtering
1
= Minimal weight filtering
2
= Moderate weight filtering
3
= Maximum weight filtering
CALIb (Calibration)
This is the calibration routine for the scale.
With display showing CALib, press the UNITS key. The display will change to show
current setting no. If scale has been previously calibrated and you wish to skip calibration
and proceed to PORT1, press the UNITS key again and internal calibration factor will be
retained.
To begin calibration, press the ZERO key to select yes and then press the UNITS key.
After pressing the UNITS key, the display will change to LOAd.
1. With LOAd displayed, press the UNITS key. The display will change to show the
currently selected scale type’s capacity. NOTE: For convenience, you can press the
ZERO key and change this to half of capacity.
2. Once the test load weight is selected, placed the test load weight on the scale platform
and then press the UNITS key.
3. Starting at the right and proceeding left, a series of dashes will appear on the display
and then starting at the left and proceeding right disappear. Next, the display will
change to the UnLOd prompt.
4. Remove all weight from the scale platform and then press the UNITS key.
5. Starting at the right and proceeding left, a series of dashes will appear on the display
and then starting at the left and proceeding right disappear. Next, the display will
change to the PORT1 prompt. Calibration is complete. Proceed to PORT1.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cardinal APS8 and is the answer not in the manual?

Cardinal APS8 Specifications

General IconGeneral
BrandCardinal
ModelAPS8
CategoryScales
LanguageEnglish

Related product manuals