8527-0544-2M Rev G ļ‚Ÿ APS Enterprise Scale 16
SECURITY SEAL INSTALLATION
If your metrology laws require the use of physical sealing, a lead and wire security seal can be
installed on the APS Enterprise scale to prevent the Calibration Switch Access screw located
on the bottom of the scale base from being removed to gain access to the calibration switch
thereby preventing unauthorized access to the calibration adjustments.
NOTE: In order to install the lead and wire security seal, two screws must be replaced with
two drilled screws. They are available from the Cardinal/Detecto Parts Department. The
part numbers are:
6021-0605 (SCW FILLISTER. MACHINE-SCW 06-32X.375)
6024-0126 (WASHER FLAT)
6021-1108 (SCW FILLISTER. MACHINE-SCW 10-32X.375)
To install a lead and wire security seal, follow the steps provided below:
1. Remove the Calibration Switch Access
screw and replace it with the drilled 10-32
screw. Make certain it is securely tighten
before proceeding.
2. Remove the PCB Cover screw as shown
and replace it with the drilled 6-32 screw
and flat washer. Make certain it is
securely tighten before proceeding.
3. Thread the sealing wire through the hole
in the drilled screws.
4. Pull the wire tight and install the lead seal.
5. The screws cannot be removed without
damaging the seal.
Calibration Switch
Access Screw
PCB Cover Screw
6-32 Screw and Flat Washer
10-32 Screw