EVENT COUNTER
The ProDoc Series Scale has been designed with an Event Counter type of security seal. When
selected, the ProDoc display will show a 3-digit number representing the Calibration and Setup
counters.
To Review the Event Counter
1. If scale is on, press ON/OFF key.
2. Display will show oFF and turn scale off.
3. Press and hold the UNITS key and then press the ON/OFF key.
4. The display will turn on all digits and show the software version for a few seconds.
5. Release all keys.
6. The display will show CALCH for two seconds and then change to show the 3 digit
calibration counter.
7. Next, the display will change to show SetCH for two seconds and then change to show
the 3 digit configuration (setup) counter.
8. To return to normal operation, press the UNITS key.
9. Otherwise, press ON/OFF key to turn off scale.
0044-M340-O1 Rev C ProDoc Series