88 
 
alarm and 12 records (1 minute) after. Each record is associated with a time stamp. These collections 
are saved as csv files. 20 files can be stored in total. 
The black box file can be uploaded with Carrier S-Proview. Once the upload is done, the original files 
saved in controller will be erased.  
5.26.  Pressure Transducer Calibration 
When pressure transducer calibration is performaned, entering value shall result in a recalculation for 
the slope and intercept of the 2 values. Thereafter, the new slope and intercept values shall be used for 
voltage ratio to pressure conversions. 
Step 1: Set CALIBRATION ENABLE for certain pressure transducer 
Step 2: Put the transducer at atmosphere and enter 0 PSIG(0 kPa) to CALIB PRESS1 
Step 3: Give the transducer a known pressure (100-250 PSI (690-1700 kPa) for XR, 10-30 PSI (70-200 
kPa) for DV) via Gauge and enter this pressure to CALIB PRESS2 
Step 4: Recalculate calibrated Slope and Intercept 
5.27.  Temperature Sensor Calibration 
The four water temperature sensors can be calibrated per the following steps: 
 
Step 1: Place the temperature sensor at 32 °F (0 °C) 
Drop each temperature sensor in a 32 °F (0 °C) water solution. 
 
Step 2: Read the sensor raw temperature in maintenance screen 
Go to maintenance screen, read the corresponding sensor raw temp 
Calculate the offset to be applied = 32 - sensor raw temp 
Example:    
ECW sensor raw temp reads 32.6 °F 
ECW temperature offset must be configured to -0.6 ^F (32–32.6 = -0.6) 
 
Step 3: Configure the sensor offset 
In temperature calibration screen, set the corresponding sensor temperature offset calculated in step 2.  
Note, the offset cannot exceed +/- 2 ^F.  
Example: ECW temperature offset must be configured to -0.6 ^F (32-32.6= -0.6) 
 
Step 4: (optional) Verify that the sensor offset is configured correctly 
Verify the following: 
- Sensor raw temp has not changed. 
- Sensor temperature offset now shows the configured value.