EasyManua.ls Logo

Carrier SmartVu control - 5.28. ISM Calibration; 5.28.1. ISM Calibration on J8 B 4-20 mA Output; 5.28.2. ISM Calibration on J6 0-10 v Input; 5.29. Alarm email function

Carrier SmartVu control
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SRDC-UG-2-14-000
REVISION
DATE
PAGE
19 Series Carrier®
SmartVu
TM
Controls IOM
2.3
Dec 4, 2020
89 OF 168
89
- Sensor control value is calibrated, and now shows 32 °F.
Example: ECW sensor raw temp” still reads 32.6 °F
ECW temperature offsetnow shows -0.6 ^F
Entering Chilled Water” now shows 32 °F.
5.28. ISM Calibration
5.28.1. ISM Calibration on J8B 4-20mA Output
Step 1: Input a test output mA value on the touchscreen UI. The controller will make ISM output this mA
value.
Step 2: Input the actual reading of mA value from VFD or mA meter
Step 3: Recalculate calibration factor
Calibration Factor = 1000 * (Actual Reading mA - output mA)
Range: 0-2000.
5.28.2. ISM Calibration on J6 0-10v Input
Step 1: When chiller is running, record the VFD feedback voltage and the actual reading on ISM.
Step 2: Shut down the chiller, enable ISM Calibration
Step 3: Input the VFD feedback voltage
Step 4: Input the actual reading of ISM
Step 5: Recalculate calibration factor
Calibration Factor = 1000 * (VFD feedback voltage / ISM actual reading voltage)
Range: 800-1200
5.29. Alarm email function
Carrier® SmartVu
TM
provides alarm email function
Once activate this function (in E-mail configuration menu) and set a valid E-mail address, and when any
alarm occurs, Carrier® SmartVu
TM
will automatically send an E-mail to the mail address to inform the
corresponding alarm information.
The corresponding unit information will be included in the E-mail, which should be input in the controller
information menu "position description" in advance.

Table of Contents

Related product manuals