CAL 1
FUNCTION: SET Maximum Capacity
Range 1 ~ 999,999 kg (CI-2001A)
1 ~ 99,999 kg (CI-2001B)
KEY
DISPLAY DESCRIPTION
: Increase of no.
: Shift of digit
ENTER(LIGHT) :
Store and
move into next menu
CAL 1 condition
100 kg/lb
10000 kg/lb
REF 1. The maximum capacity means the maximum weight that scale can measure.
CAL 2
FUNCTION: SET Minimum Division
Range 0.0005 ~ 100 kg/lb
KEY
DISPLAY DESCRIPTION
: Input the next
division.
ENTER(LIGHT) :
Store and
move into next menu
CAL 2 condition
1 kg
0.01 kg
REF 1. The minimum division means the value of one division.
REF 2. External resolution is obtained by division the min. division by the
maximum capacity. Set the resolution to be within 1/30,000.