3
4
ERROR MESSAGE
(1) ERRORS IN WEIGHTING MODE
Err 02
Reason
Load cell connection failure or error in A/D conversion part.
Trouble shooting
Check the load cell connector to see if the polarity of signal is reversed.
Err 06
Reason
Error in printer connection
Trouble shooting
Check with printer connector
If there is no problem with printer and printer connector, please request
A/S to head office.
Err 08
Reason
The ZERO key or TARE key is adjusted not to be operated under the unstable
condition.
Trouble shooting
Press Zero or Tare key in stable condition
Err 09
Reason
Current weight deviates from zero range.
Trouble shooting
Press the ZERO key within 10% of the maximum capacity.