EasyManua.ls Logo

CAS CI-2001A - Error Messages in Weighting Mode; Err 02: Loadcell;A;D Error; Err 06: Printer Connection Error; Err 08: Unstable Key Operation

CAS CI-2001A
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
4
ERROR MESSAGE
(1) ERRORS IN WEIGHTING MODE
Err 02
▣ Reason
Load cell connection failure or error in A/D conversion part.
☞ Trouble shooting
Check the load cell connector to see if the polarity of signal is reversed.
Err 06
▣ Reason
Error in printer connection
☞ Trouble shooting
Check with printer connector
If there is no problem with printer and printer connector, please request
A/S to head office.
Err 08
▣ Reason
The ZERO key or TARE key is adjusted not to be operated under the unstable
condition.
☞ Trouble shooting
Press Zero or Tare key in stable condition
Err 09
▣ Reason
Current weight deviates from zero range.
☞ Trouble shooting
Press the ZERO key within 10% of the maximum capacity.

Table of Contents