23
CAL 3
FUNCTION: Setting Weight In Span CALIBRATION
Range 1 ~ 999,999 kg (CI-2001A)
1 ~ 99,999 kg (CI-2001B)
KEY
DISPLAY DESCRIPTION
: Increase of no.
: Shift of digit
ENTER(LIGHT) :
Store and
move into next menu
CAL 3 condition
100 kg
10000 kg
REF 1. The weight shall be within the range of 10 % ~100 % of maximum weight.
REF 2. If the Setting Weight is under the 10% of the Maximum Capacity,
Error message ("ERR 22") will occur.
REF 3. If the Setting Weight over the Maximum Capacity,
Error message ("ERR 23") will occur.
CAL 4
FUNCTION: Zero Calibration
KEY
DISPLAY DESCRIPTION
ENTER(LIGHT) :
Zero calibration and
Move into next menu.
CAL 4 condition
Unload the tray and press ENTER
Under zero calibration
Zero calibration is completed.
REF 1. If Zero calibration is done without any error, GOOD message is displayed
and program moves into CAL 5 automatically.
REF 2. If the "ZERO" key is press, only Zero Calibration is completed and
program moves SAVE & EXIT mode. Press ENTER Key.