29
FUNCTION Digital filter
DISPLAY DESCRIPTION
F6-01 high speed, weak vibration
F6-09 low speed, strong vibration
F06
SET
(01
09)
FUNCTION Device ID(Identification number of each indicator)
DISPLAY DESCRIPTION
F12-01 Device ID 01
F12-99 Device ID 99
F12
SET
(00
09)
FUNCTION RS232C baud rate set
DISPLAY DESCRIPTION
F10-00 2400
F10-01 4800
F10-02 9600
F10-03 19200
F10
SET
(00
03)
RS232C COMMUNICATION
FUNCTION Output method RS232C data
DISPLAY DESCRIPTION
F11-00 No data output
F11-01 Output data in state of stable & unstable
F11-02 Output data only in stable state
F11-03 Output data when data is required
F11
SET
(00
03)
Note 1. The default value is set to 0 at factory.
Note 2. It is available in case of setting 0 in F13.
Note 3. If device ID(F12) is received in case of setting 3 in F11, weight data will be
transmitted.