50
ERROR MESSAGE & TROUBLE-SHOOTING
1. In Weighing Mode
CH 01
Reason: The calibration isn’t executed or the data of internal memory is lost by
a electrical shock.
Trouble shooting: Perform the required specification again in calibration mode.
CH 02
Reason: The load cell is connected wrongly or A/D conversion part has a fault.
Trouble shooting: Check that the load cell is well connected with main unit.
CH 03
Reason: The indicator can’t be initialized due to platform shaking.
Trouble shooting: Put the platform on a flat place without shaking and power
on.
CH 04
Reason: The initial zero range exceeds above 10% of maximum capacity.
Trouble shooting: Check the platform.
CH 05
Reason: Pressing any key for long time or having problems in key part.
Trouble shooting: If it can’t be found problems in key part, please call A/S.
noprt
Reason: When it is setting F13=0 or when you press print-key to print current
and unit weight at count mode.
Trouble shooting: If you want to print, do it in setting F13=1.
And don’t print unit weight, current weight at count mode.
Over
Reason: The gross weight is over maximum capacity.
Trouble shooting: Don’t load the article whose weight is heavier than the
maximum capacity.
CH 15
Reason: The display value is over 99999.